This section clarifies that ( especially in the life sciences) image sensors or cameras themselves are not the only possibility for the generation of images. I discuss ultrasound imaging devices, computer tomography and its derivatives, and magnetic resonance imaging (MRI).
Ultrasound Imagers
Ultrasound imaging devices are based on the principle of the different sound impedances of different body tissue classes, resulting in reflection and refraction of the sound waves, similar to the respective optical effects. An ultrasound head is used for the emission as well as for the reception of short sound packets, as shown in Figure 2.36.
Figure 2.36. Principle of Ultrasound A and M Mode [19]
The A mode simply displays the reflections of the tissue boundaries (Figure 2.36). This information plotted over time is called M mode . Figure 2.36 also shows an M mode display in connection with the electrocardiogram (ECG) [19].
For ultrasound "images" to be obtained from the area of interest, it is necessary to move the sound head over a defined region (2D ultrasound or B mode ). Sound heads that perform this movement automatically are called ultrasound scanners . Mechanical scanners, used in the early days of ultrasound imaging devices, are nowadays replaced by electronic scanning heads. The example shown in Figure 2.37 uses a curved array of ultrasound elements, which are activated in groups and electronically shifted in the desired direction.
Figure 2.37. Curved Array Ultrasound Head and Corresponding Image
Some more information about sound reflection and refraction: Like the corresponding optical effects, the reflection of sound waves at tissue boundaries with different sound impedance is described by the reflection factor R :
Equation 2.22
with Z i as the sound impedance of the respective area and the transmission factor T :
Equation 2.23
This is why a special ultrasound transmission gel has to be applied to the sound head before it is attached to the skin; otherwise , almost all of the sound energy would be reflected. Table 2.6 shows examples for the reflection factor R at some special tissue boundaries.
Table 2.6. Typical Values of the US Reflection Factor R
from/to: |
Fat |
Muscle |
Liver |
Blood |
Bone |
---|---|---|---|---|---|
Water: |
0.047 |
0.020 |
0.035 |
0.007 |
0.570 |
Fat: |
0.067 |
0.049 |
0.047 |
0.610 |
|
Muscle: |
0.015 |
0.020 |
0.560 |
||
Liver: |
0.028 |
0.550 |
|||
Blood: |
0.570 |
Computed Tomography
The principle of computed tomography (CT) is based on x-ray imaging, which makes use of the characteristic attenuation of x-rays by different tissue types. X-ray imaging generates only a single image, showing a simple projection of the interesting region. CT, on the other hand, can display cross sections of the body, so the imaging device has to "shoot" a large number of attenuation profiles from different positions around the body. An important issue is the reconstruction of the resulting image.
Looking at the simplest case of a homogeneous object of interest, [8] we see that the intensity I of the detected radiation decreases exponentially with the object width d and is therefore given by
[8] This case also includes radiation of one single frequency.
Equation 2.24
whereas the attenuation P is defined by the product of the attenuation coefficient m and the object width d :
Equation 2.25
finally, the attenuation coefficient itself is therefore
Equation 2.26
CT Device Generations
Computed tomography devices are usually classified in "generations," as shown in Figure 2.38:
Figure 2.38. CT Device Generations 1 to 4 [20]
CT Image Calculation and Back Projection
As you can imagine, it is not easy to obtain the resulting image (the spatial information), because each beam [9] delivers the total attenuation value of its route; we defined the resulting intensity I in Eq. (2.24) for the simple case of a homogeneous object. In most cases, this will not be true; Eq. (2.24) therefore becomes
[9] Here, we are considering a CT device of generation 1.
Equation 2.27
with m i as the attenuation coefficient in a specific area and d i as the respective length of this area, for the intensity I in a certain direction.
Figure 2.39. Simple Calculation of a CT Image [20]
Special algorithms are necessary for computation of the original values m i of a certain area. The following solutions can be used:
A good example, which illustrates the method of back projection, can be found at the National Instruments web page www.ni.com; just go to "Developer Zone," search for "tomography," and download the LabVIEW library Tomography.llb (see Figure 2.40).
Figure 2.40. Tomography Simulation (National Instruments Example)
Figure 2.41. Iterative Calculation of a CT Image
Exercise 2.5: Iterative Calculation of a CT Image.
Try to make the iterative calculation (method 2) in a LabVIEW VI (front panel shown in Figure 2.41). Here, the intensity values of four directions are used, starting with the one to the right. The first estimation is based on the equal distribution of the intensity values over the entire lines. The values of the next direction (bottom right) are calculated and compared to the values of the original image. The difference of these results is applied to the first estimation, resulting in the second estimation, and so on.
As you will find out, the values already are quite close to the original values. You can get the values closer to the original if you run the previously described steps again with the new values. After a third run, the differences will be too small to notice.
Single Photon Emission Computed Tomography (SPECT)
This method is quite similar to computer tomography with respect to image generation. The main difference is that the total emitted gamma radiation, not an attenuation coefficient, of a specific direction is measured. Therefore, the same imaging techniques (usually filtered back projection) can be used.
Positron Emission Tomography (PET)
PET is based on the detection of gamma quants that result from the collision of an electron with a positron. To make this possible, interesting molecules have to be marked before imaging with a positron radiator; we do not discuss this process here.
In general, for PET images the same image generation processes that have been described above can be used; here, the main difference is that the machine cannot decide ahead of time which direction is to be measured. However, it is possible to afterwards assign specific events to the appropriate projection directions.
Magnetic Resonance Imaging
Magnetic resonance imaging (MRI; sometimes NMRI for nuclear MRI) uses a special property of particular atoms and especially their nuclei; this property, called spin , is related to the number of protons and neutrons. The spin is represented by the symbol I and always occurs in multiples of 1/2. For example, the nucleus of 1 H has a spin of I = 1/2 and the nucleus of 17 O a spin of I = 5/2.
In Figure 2.42, the spin of a nucleus is visualized by an arrow (up or down), which indicates that the nuclei are "rotating" around the arrow axis clockwise or counterclockwise, respectively. If an external magnetic field B is applied, the (usually unoriented) spins may align parallel or anti-parallel to this field.
Figure 2.42. Separation of Energy Levels According to Spin Directions
The energy necessary for the transition from the parallel to the anti-parallel state depends on the value of the external magnetic field B and is given as
Equation 2.28
with h as Planck's constant and g as the intrinsic gyromagnetic ratio. Integrated in Eq. (2.28) is the Larmor equation
Equation 2.29
which indicates that the frequency related to the nuclei [10] depends on the value of the external magnetic field.
[10] This is the so-called precession frequency.
If the energy of Eq. (2.28) is applied to some nuclei, a large number of them are raised from their state of lower energy to the state of higher energy. After a certain time, the nuclei return to their original state, emitting the energy D E .
Figure 2.43. Relaxation Times T 1 and T 2
The magnetization M can be split into a component M xy residing in the xy plane and a component M z in z direction. After excitation of the nuclei with a certain frequency pulse, both magnetization components follow the equations
Equation 2.30
and
Equation 2.31
(see also Figure 2.43) with the specific relaxation times T 1 and T 2 . Table 2.7 shows some different values for some important tissue types.
If the Larmor equation (2.29) can be modified to
Equation 2.32
with w ( r i ) as the frequency of a nucleus at the location r i and describing the gradient location-dependent amplitude and direction, the location r i of the nucleus can be specified. The final imaging techniques are quite similar to those used in computer tomography but offer a number of additional possibilities:
Table 2.7. Typical Values of Relaxation Times T 1 and T 2 at 1 Tesla
Tissue Type |
Relaxation Time T 1 |
Relaxation Time T 2 |
---|---|---|
Muscle: |
730 ± 130 ms |
47 ± 13 ms |
Heart: |
750 ± 120 ms |
57 ± 16 ms |
Liver: |
420 ± 90 ms |
43 ± 14 ms |
Kidney: |
590 ± 160 ms |
58 ± 24 ms |
Spleen: |
680 ± 190 ms |
62 ± 27 ms |
Fat: |
240 ± 70 ms |
84 ± 36 ms |
Figure 2.44. MRI Images: Based on T 1 (left) and T 2 (right) of a Knee Joint
Image Distribution |
Introduction and Definitions
Image Acquisition
Image Distribution
Image Processing
Image Analysis
About the CD-ROM