A Nonuniform Detector Aperture for CT

logo Website of Bruno R. Preiss
next up contents external
Next : On the Transport of Up: Journal Articles Previous: Short Packet Transfer Performance

  R.  M. Henkelman and Bruno  Richard Preiss. Journal of Computer Assisted Tomography , 5(3):401-408, June 1981[9].
The finite width of the X-ray beam determined by the source and detector aperture limits the resolution of computed tomography (CT) images. Improved resolution can be obtained by reducing the width of the detector, but at a cost in dose utilization efficiency. We show that it is possible to improve the resolution of a CT image with a minimum cost in dose efficiency by varying the sensitivity (apodizing) the detector response function. Further, we demonstrate how to construct detectors with nonuniform response by changing the reflective coating on the scintillator. Finally, we show improved calculated point response functions for such modified detectors.

Copyright 1981 by Raven Press, New York.

external Full text. external BibTeX entry.



bruno Copyright 2002 by Bruno R. Preiss, P.Eng. All rights reserved.